Vaisala HUMICAP HMP41 User's Guide Page 22

  • Download
  • Add to my manuals
  • Print
  • Page
    / 96
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 21
User's Guide ______________________________________________________________________
20 __________________________________________________________________ M211280EN-A
Duct Installation Kit
The duct installation kit includes a plastic pipe with a flange (Vaisala
order code: 215619). To install the HMP110 probe with the duct
installation kit, drill a hole to the duct wall, assemble the probe to the
duct installation kit, slide the probe head through the hole, and attach the
flange to the duct wall with four screws. See section Probe Assembly
with Duct Installation Kit on page 21 for details.
24
L
1
0505-176
Figure 5 Probe Installation with the Duct Installation Kit
The following explanations refer to Figure 5 above:
1 = Tension screw
Distance L can be adjusted and locked in place with the tension screw.