Set/Show Analog Output Error Levels
Use the AERR command to show or set the analog output error levels.
AERR [level1 level2]<cr>
where
level1 level2
= Error levels for analog output channel
Examples:
>aerr
Ch1 Error level : 0.0
Ch2 Error level : 0.0
>
>aerr 10.8 10.8
Ch1 Error level : 10.8
Ch2 Error level : 10.8
>
Set/Show Analog Output Quantity and
Scaling
After you have defined the quantities with the CALCS command, use the
ASEL command to show or set the quantity output by the analog
channels, and the scaling that is used. If you do not enter scaling limits,
default values are used. Valid quantities are: RH, T, TD, TDF, A, X, H,
TW, PWS, PW and NONE. NONE parameter disables analog output and
output is forced to error level.
ASEL [q1 q2] [lo1 hi1 lo2 hi2]<cr>
where
q1 q2
= Analog output channel quantities
lo1 lo2
= Low limits of scaling
hi1 hi2
= High limits of scaling
Syntax for single-parameter transmitters:
ASEL [q] [lo hi]<cr>
Where
q
= Analog output channel quantity
lo
= Low limit of scaling
hi
= High limit of scaling