Vaisala HUMICAP HMP41 User's Guide Page 34

  • Download
  • Add to my manuals
  • Print
  • Page
    / 96
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 33
User's Guide ______________________________________________________________________
32 __________________________________________________________________ M211280EN-A
NOTE
Single-parameter transmitters HMT130H and HMT130T have only one
quantity (T for HMT130T, and RH, TD, TDF, A, X, H, TW, PWS, PW
for HMT130H.)
Set/Show Calibration Date
Use the CDATE command to show or set the calibration date field.
(Calibration refers to the calibration of the HMT130's analog outputs.)
CDATE [date]<cr>
where
date
= Calibration date in format yyyy-mm-dd
Examples:
>cdate
Calibrated : "NOT SET"
>
>cdate 2009-06-22
Calibrated : "2009-06-22"
>
Set/Show Calibration Information
Use the CTEXT command to show or set the calibration info field.
(Calibration refers to the calibration of the HMT130's analog outputs.)
CTEXT [location]<cr>
where
location
= Freely defined text string typically indicating the
calibration location
Examples:
>ctext
Calib. info: "VAISALA HELSINKI"
>
>ctext "Calibration lab 2"
Calib. info: "Calibration lab 2"
>
Page view 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 95 96

Comments to this Manuals

No comments