Performance
BAROMETRIC PRESSURE
Pressure range 500 ... 1100 hPa, 50 ... 1100 hPa
Accuracy 500 ... 1100 hPa 500 ... 1100 hPa 50 ... 1100 hPa
CLASS A CLASS B
Linearity ±0.05 hPa ±0.10 hPa ±0.20 hPa
Hysteresis* ±0.03 hPa ±0.03 hPa ±0.08 hPa
Repeatability* ±0.03 hPa ±0.03 hPa ±0.08 hPa
Calibration
uncertainty**
±0.07 hPa ±0.15 hPa ±0.20 hPa
Accuracy at +20 °C*** ±0.10 hPa ±0.20 hPa ±0.30 hPa
Temperature
dependence****
±0.1 hPa ±0.1 hPa ±0.3 hPa
Total accuracy
(-40 ... +60 °C/
-40 ... +140 °F)
±0.15 hPa ±0.25 hPa ±0.45 hPa
Long-term stability/year ±0.1 hPa ±0.1 hPa ±0.2 hPa
Response time (100 % response)
one sensor 2 s
•
1 s
•
1 s
•
Pressure units
hPa, mbar, kPa, Pa, inHg, mmH20, mmHg, torr, psia
* Defined as ±2 standard deviation limits of endpoint non-linearity,
hysteresis error or repeatability error and calibration.
** Defined as ±2 standard deviation limits of accuracy of the working
standard including traceability to NIST.
*** Defined as the root sum of the squares (RSS) of endpoint non-linearity,
hysteresis error, repeatability error and calibration uncertainty at room
temperature.
**** Defined as ±2 standard deviation limits of temperature dependence over
the operating temperature range.
RELATIVE HUMIDITY
Measurement range 0 ... 100 % RH
Accuracy (including non-linearity,
hysteresis, and repeatability at
+15 ... +25 °C ±1 %RH (0 ... 90 % RH)
±1.7 %RH (90 ... 100 %RH)
at -20 ... +40 °C ±(1.0 + 0.008 x reading) %RH
at -40 ... +60 °C ±(1.5 + 0.015 x reading) %RH
Factory calibration uncertainty (+20 °C )
(Defined as ±2 standard deviation
limits. Small variations possible,
see also calibration certificate.)
± 0.6 % RH (0 ... 40 %RH)
± 1.0 % RH (40 ... 97 %RH)
Sensor
for typical applications Vaisala HUMICAP
®
180 or 180R*
for applications with chemical
purge/warmed probe Vaisala HUMICAP
®
180C or 180RC*
Response time (90 %) at +20 °C (+68 °F) in still air
with grid filter 8 s / 17 s*
with grid + steel netting filter 20 s / 50 s*
with sintered filter 40 s / 60 s*
* with HUMICAP
®
180R or 180RC sensor
TEMPERATURE
Measurement range, all probes -40 ... +60 °C (-40 ... +140 °F)
Accuracy at +20 °C (+68 °F) ± 0.2 °C (± 0.4 °F)
Temperature units °C, °F
Temperature sensor PT100 RTD 1/3 Class B IEC 751
ACCURACY OVER TEMPERATURE RANGE
Operating Environment
Operating temperature -40 ... +60 °C (-40 ... +140 °F)
with display 0 ... +60 °C (+32 ... +140 °F)
Humidity range non-condensing
Electromagnetic compatibility EN61326-1:1997 + Am1:1998
+Am2:2001; Industrial Environment
Δ
°C
°C
0.5
0.4
0.3
0.2
0.1
-0.1
-0.2
-0.3
-0.4
-0.5
0
0-40 -20 20 40 60
Technical Data
The display also shows the WMO pressure
trend ∆P 3h and tendency of 0 ... 9.
Inputs and Outputs
Operating voltage 10 ... 35 VDC, 24 VAC
with optional power supply module 100 ... 240 VAC, 50/60 Hz
Power consumption at +20 °C (U
in
24 VDC)
RS-232 max. 28 mA
U
out
3 x 0 ... 1 V/0 ... 5 V/0 ... 10 V max. 33 mA
I
out
3 x 0 ... 20 mA max. 63 mA
display and backlight +20 mA
during chemical purge max. +110 mA
during probe heating +120 mA
Settling time at power-up (one sensor)
class A 4 s
class B 3 s
External loads
current outputs R
L
< 500 ohm
0 ... 1 V output R
L
> 2 kohm
0 ... 5 V and 0 ... 10 V outputs R
L
> 10 kohm