APPENDIX 1 SERIAL COMMANDS __________________________________________________________
94 _____________________________________________________________________ U263EN-1.4
Inactivating the time output
>ftime off
Form. time : OFF
>r
RH= 19.4 %RH T= 26.1 ’C
RH= 19.8 %RH T= 26.1 ’C
RH= 20.6 %RH T= 26.1 ’C
RH= 20.5 %RH T= 26.1 ’C
RH= 19.9 %RH T= 26.1 ’C
RH= 19.6 %RH T= 26.1 ’C
...
FDATE Adding date to output
FDATE xxx <cr>
xxx = ON or OFF
When FDATE is activated, the current date is output at the beginning of the
output line. The date is set with command DATE. After RESET or power on
the current date is 1991-01-01.
Activating the date output
>fdate on
Form. date : ON
>r
1995-03-10 RH= 21.1 %RH T= 26.0 ’C
1995-03-10 RH= 21.3 %RH T= 26.0 ’C
1995-03-10 RH= 23.1 %RH T= 26.0 ’C
1995-03-10 RH= 22.2 %RH T= 26.0 ’C
1995-03-10 RH= 20.6 %RH T= 26.0 ’C
...
Inactivating the date output
>fdate off
Form. date : OFF
>r
RH= 20.2 %RH T= 26.0 ’C
RH= 19.9 %RH T= 26.0 ’C
RH= 19.8 %RH T= 26.0 ’C
RH= 19.7 %RH T= 26.0 ’C
RH= 19.7 %RH T= 26.0 ’C
...