CHAPTER 3_______________________________________________________________INSTALLATION
VAISALA __________________________________________________________________________ 9
NOTE
Take care not to damage the pipe of the probe. If the pipe is damaged,
the probe head is less tight and it will not go through the clasp nut.
Mounting; overview
>10.5mm
fitting body
hex = 24mm
tapered thread
R1/2 ISO 7/1
ø19mm drilling
sealing with:
1. LOCTITE® No 542 + activ. No 7649 (t=-55...+150 °C)
2. MEGA-PIPE EXTRA No 7188 (t=-55...+170 °C)
3. PTFE tape (t=-60...+210 °C) NOTE: the tape does not lock
the parts together. Therefore, use two fork spanners (hex 24 and
27 mm) for tightening and opening the clasp nut of the probe
parallel thread
G1/2 ISO 228/1
(BS 2779, JIS B0202)
>40mm
Process or pipe wall
FIGURE 3-3 Sealing and thread cutting for the fitting body.
The fitting body can be installed e.g. on standard pipe fittings (G 1/2
ISO 228/1) or on a thread in the process wall. If the wall thickness is
less than 10.5 mm, it is recommended to use a welded sleeve (see
FIGURE 3-3). Note that the minimum recommended distance of the
fitting body and probe head is 40 mm (see FIGURE 3-3).
Adjust the probe to a suitable distance according to the type of
installation, and tighten the clasp nut first manually. Then, mark the
fitting body and the clasp nut and tighten the nut a further 50...60°
with a fork spanner (see FIGURE 3-4).
60°
probe
fitting body
clasp nut
max.
a pen
FIGURE 3-4 Tightening the clasp nut.