• Enter the measured lower end current/voltage on channel 1.
• Enter the measured upper end current/voltage on channel 1.
• Connect the meter to the output of channel 2 and enter the
measured lower end current/voltage on channel 2.
• Enter the measured upper end current/voltage on channel 2.
Using LED commands
If both the analogue outputs and humidity/temperature channels are
calibrated, the analogue outputs should be calibrated first. This applies
only when the calibrations are done using the LED commands!
• connect an ammeter/voltmeter to the analogue outputs (connector
X2)
Disconnect the security lock jumper!
• Give command ll.
• the LED on the left flashes; set the low end of channel 1 with the
arrow keys and press ENT
• the second LED from the left flashes; set the high end of channel 1
with the arrow keys and press ENT