CHAPTER 3_______________________________________________________________INSTALLATION
VAISALA _________________________________________________________________________ 13
STEP 1: mount the probe with the ball valve assembly closed;
tighten the clasp nut manually.
fitting body
R1/2 cone, seale
d
bushing
R1/2 cone
sealed
bushing R1/2 cone/G1/2(40 bar)
e.g. Camozzi 2520-1/2-1/2
ball valve 1/2" (40 bar)
e.g. Atlas Copco:BAL-1A 15 (G1/2)
nipple
R1/2 cone
sealed
(the bushing serves for
moving the probe (sinter)
to such a distance from the
ball valve that the valve
can be closed)
>30 mm
clasp nut
FIGURE 3-8 Installing the probe through the ball valve assembly;
step 1.