Vaisala DMP248 Service Manual Page 127

  • Download
  • Add to my manuals
  • Print
  • Page
    / 145
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 126
APPENDIX 4 ________________________________________________________________________
VAISALA ________________________________________________________________________ 121
Set the serial port of the terminal to 4800 baud, even
parity, seven data bits and one stop bit, full duplex (4800 E 7 1 FDX).
The serial settings of the transmitter must also be 4800 E 7
1 FDX and the transmitter must be in STOP mode. If these factory
settings have been changed, they must be changed back. Connect the
RS 232C port of the terminal to connector X17 on the top of the main
board and switch the power on.
RX GND TX
X17
Single loop wiring
Set the address of the transmitter; it can be any number
between 1 and 99. In this example the address is 22:
>addr 22
Address : 22
Set the serial bus settings according to your network
specifications. This setting will become valid after next RESET or
power off:
>seri 2400 e 7 1 h
2400 E 7 1 HDX
Set the transmitter in POLL mode:
>smode poll
Serial mode : POLL
NOTE
The SMODE command must be given last.
NOTE
The transmitter outputs no prompt (>) after the SMODE POLL
command and it only reacts to commands which include its address.