Vaisala DMP248 Service Manual Page 32

  • Download
  • Add to my manuals
  • Print
  • Page
    / 145
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 31
OPERATING MANUAL_______________________________________________________________
26 _____________________________________________________________________ U263EN-1.4
Selecting the analogue outputs
The DMP248 transmitter can be ordered ready with the current or
voltage outputs required. If the outputs need to be changed, move the
jumpers in connector X15 into positions as shown in FIGURE 4-2.
CH2 CURRENT OUTPUT
CH1 0 ... 1 VOLTAGE OUTPUT
OPENED COVER OF THE DMP248
CH1
CH2
X15
CURRENT OUTPUTS
0 ... 20 / 4 ... 20 mA
VOLTAGE OUTPUTS
0 ... 5 V / 0 ... 10 V
VOLTAGE OUTPUTS
0 ... 1 V
CH1
CH2
CH1
CH2
CH1
CH2
CH1
CH2
CH2
CH2
CH2
CH1
CH1
CH1
FIGURE 4-2 Selecting the analogue outputs with jumpers.
The software also has to be informed which outputs are in use. This is
done either through the serial interface or the menus on local display
when in use. The serial command is AMODE and the display/keypad
command 0RGHÕ$QDORJRXWSXWVÕ0RGH(see Chapter 5
commands). If the outputs need to be scaled, see serial
Page view 31
1 2 ... 27 28 29 30 31 32 33 34 35 36 37 ... 144 145

Comments to this Manuals

No comments