CHAPTER 6_______________________________________________________________ CALIBRATION
VAISALA _________________________________________________________________________ 55
CHAPTER 6 CALIBRATION
The DMP248 transmitter has been fully calibrated at the factory so
there should be no immediate need for recalibration. The transmitter
should be calibrated only if there is reason to believe that the
adjustments have changed. The adjustments of the temperature
measurement channel and the analogue outputs are particularly stable
and in normal circumstances, there is no need to recalibrate them.
Humidity calibration should be performed at least once a year.
Humidity calibration
The DMP248 transmitter is calibrated against two accurate RH
references. The calibration can be performed by the end-user, or the
instrument can be sent to Vaisala or a Vaisala representative. A two-
point calibration can be performed with Vaisala’s HMK15 or
HMK13B Calibrator.
A Ø 13.5 adapter must be used when calibrating with the HMK13B
Calibrator. The adapter (part no. 16611) can be ordered from Vaisala
or Vaisala representatives.
Calibration can be performed by giving the commands using the press
switches inside the housing, through the serial bus (serial commands)
or through the menus on the local display (display/keypad
commands).
When LED commands are used and when the two analogue channels
do not output either relative humidity and/or temperature, relative
humidity is calibrated on channel 1 and temperature is calibrated on
channel 2. The calibration ranges are 0...100 %RH and -20...+80 °C.
When the transmitter is calibrated at two points, the points must be
either 50 %RH or 50 C apart from each other.