Vaisala DMP248 Service Manual Page 56

  • Download
  • Add to my manuals
  • Print
  • Page
    / 145
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 55
OPERATING MANUAL_______________________________________________________________
50 _____________________________________________________________________ U263EN-1.4
xxx = channel 1’s quantity
yyy = channel 2’s quantity (RH, T, Td, ppm)
Example: dewpoint temperature selected on channel 1 and
temperature on channel 2
>ASEL Td T <cr>
Ch1 ( Td ) lo -50.00 ’C ? -40<cr>
Ch1 ( Td ) hi 10.00 ’C ? 0<cr>
Ch2 ( T ) lo 10.00 ’C ? 0<cr>
Ch2 ( T ) hi 100.00 ’C ? 60<cr>
>
Scaling the analogue outputs
Disconnect the security lock jumper!
ASCL <cr>
Example: dewpoint is scaled in the range of -40...0 °C and tem-
perature in the range of 0...+60 °C
>ASCL <cr>
Ch1 ( Td ) lo -50.00 ’C ? -40<cr>
Ch1 ( Td ) hi 10.00 ’C ? 0<cr>
Ch2 ( T ) lo 10.00 ’C ? 0<cr>
Ch2 ( T ) hi 100.00 ’C ? 60<cr>
>
Output via the serial bus
Starting the measurement output
R <cr>
Starts output of measurements to the peripheral devices (RUN mode);
the only command that can be used is S (stop).
The output format can be changed with command FORM (see
Appendix 1).
Page view 55
1 2 ... 51 52 53 54 55 56 57 58 59 60 61 ... 144 145

Comments to this Manuals

No comments