APPENDIX 4 ________________________________________________________________________
124 ____________________________________________________________________ U263EN-1.4
Checking the serial port network operation
Normally, measurement readings are asked when the transmitter is in
POLL mode; then the command must include the address of the
transmitter. If the settings of the transmitter need to be changed, the
transmitter is switched to STOP mode with command OPEN;
commands can then be given without address. When the line to the
transmitter is closed, it returns to POLL mode.
STOP mode
Open the line to the transmitter:
open 22<cr>
DMP 22 line opened for operator commands
Transmitter no. 22 is now temporarily set to STOP mode; it accepts
commands without address until CLOSE command is given.
Individual settings can now be easily modified. Do not open more
than one line at a time.
Use command ? to find out the settings of the active transmitter:
? <cr>
DMP 248 /1.01
CPU serial nr : 0
Keyboard type : 0
Address : 7
Output units : metric
Baud P D S : 4800 E 7 1 FDX
Serial mode : STOP
Output intrv. : 0 min
Mtim : 32
Pressure : 1013.25
Analog outputs
Ch1 0.00 ... 10.00 V
Ch2 0.00 ... 10.00 V
Ch1 ( RH ) lo 0.000 %RH
Ch1 ( RH ) hi 100.000 %RH
Ch2 ( T ) lo -20.000 ’C
Ch2 ( T ) hi 180.000 ’C
Transducer :
PRB serial nr : 0
Calibr. date : 0