APPENDIX 1 SERIAL COMMANDS __________________________________________________________
84 _____________________________________________________________________ U263EN-1.4
DCAL Forced auto-calibration
DCAL<cr>
>dcal
Calibration...
any key to abort
> (appears when the auto-calibration
is completed, max. 60 - 70 seconds)
If you press any key, the calibration is interrupted. If no key is
pressed, the calibration takes place immediately. The text above is
displayed during the calibration. If the process is unstable (see
Chapter ) or relative humidity is over 10%, the calibration is not
performed. The text above is only shortly displayed and the display
returns to the measuring mode. The calibration is not performed again
even if the maximum correction is exceeded.
Analogue output commands
AMODE Setting the analogue outputs
Disconnect the security lock jumper!
AMODE a bb.bbb cc.ccc d ee.eee ff.fff <cr>
a = channel 1: U = voltage output
I = current output
bb.bbb = lower limit of channel 1
cc.ccc = upper limit of channel 1
d = channel 2: U = voltage output
I = current output
ee.eee = lower limit of channel 2
ff.fff = upper limit of channel 2
The bb.bbb, cc.ccc, ee.eee and ff.fff parameters are entered in volts or
milliamperes.
Sets the analogue outputs on channels 1 and 2. An example of this is
when the voltage output on channel 1 is set to be 0...1 V and channel 2
set to 2...10 V: