CHAPTER 3_______________________________________________________________INSTALLATION
VAISALA __________________________________________________________________________ 7
CHAPTER 3 INSTALLATION
Selecting the place of installation
Select a place which gives a true picture of the environment or
process; also select a place that is as clean as possible. Air should
circulate freely around the sensor.
It is recommended that the sensor head is installed directly in the
process through the ball valve assembly. When the ball valve
assembly is used, the chamber or the duct does not have to be emptied
or shut down for installation or removal of the probe. Install the sensor
head transversely against the direction of the process flow.
If the probe head has to be installed aside of the process gas flow or
the process is very hot or particularly dirty, the probe can be installed
in a “leak-through” position. In this installation, the probe is mounted
behind the ball valve assembly and if necessary, a cooling coil and/or
a filter can be mounted in between. The flow passes through the
sensor head and leaks out through a vent hole in the fitting body
enabling a reasonable response time. In hot and dirty processes, a
sample system can also be used.
Mounting the transmitter
In FIGURE 3-1 and FIGURE 3-2, you can see the dimensions of the
DMP248 transmitter: