CHAPTER 5________________________________________________________________ COMMANDS
VAISALA _________________________________________________________________________ 41
• Select the parameter to be changed with the arrow keys and ENT
key.
Selecting baud rate:
Selecting parity:
Selecting data bits:
Selecting stop bits:
Full duplex/half duplex:
The processor does not allow the following combinations:
• no parity, 7 data bits, 1 stop bit: if this combination is given the
DMP248 programme will change the number of stop bits to 2
• even or odd parity, 8 data bits, 2 stop bits: if this combination is
given the programme changes the number of stop bits to 1
NOTE
The serial bus settings become effective only after reset.