APPENDIX 1 SERIAL COMMANDS __________________________________________________________
92 _____________________________________________________________________ U263EN-1.4
When this is changed into 10 minutes, the command is:
>INTV 10 <cr>
Output intrv. : 10 min
The unit is changed into seconds with:
>INTV S <cr>
Output intrv. : 10 s
The change can also be done with one command:
>INTV 10 S <cr>
Output intrv. : 10 s
FORM Setting the output format
FORM <cr>
"xxx...xxx"
? zzz...zzz <cr>
xxx...xxx = old format
zzz...zzz = new format
The FORM command sets the format of the outputs generated in RUN state
and by SEND command.
NOTE
Please note that capital and small letters have different meanings.
\DD..DD\ dewpoint temperature
\TT..TT\ temperature
\UU..UU\ relative humidity
\MM..MM\ ppm concentration
\PP..PP\ absolute pressure (hPa) (manually feeded value)
\uu..uu\ unit according to the preceding variable
\n line feed <lf>
\r carriage return <cr>
\t horizontal tabulation <ht> or <tab>
\\ \
For example:
format output
\UUU.UU\ \+TT.TT\\r 100.00 +99.99 <cr>
\TTT.T\ \uu\\r\n 15.2 ’C <cr><lf>
\UUU.U\ \uuu\\+DD.D\ \uu\\r 46.9 %RH +10.8 ’C <cr>