User's Guide _______________________________________________________________________
74 ___________________________________________________________________ M210243en-A
0201-034
Figure 65 Aligning the Wind Vane
Using Compass and Reference Point
With the MAWS running, monitor the instant wind speed in the
reports sent through serial line.
1. The wind sensor cable must be connected both to the sensor and
to the Wind connector.
2. The mounting piece (2) must be placed on top of the upper tube
and the sensor must be attached to the mounting piece with the
plastic collar (3).
3. Choose a known wind direction reference point on the horizon
with the help of a compass.
4. Point the nose of the vane at the reference point.
5. Hold the vane in position and slowly rotate the mounting piece
until wind direction shows proper value.
6. Secure the mounting piece to the mast by tightening the
mounting screw (4).